| Operating Conditions | 0 - 90% RH non-condensing, 0 - 8000 ft. above sea level ^^ | ||
| Shipping/Storage Conditions | -30 to 50°C, 0 - 90% RH non-condensing | ||
| Protection from Water Ingress | Ordinary | ||
| Protection from Flammable Gases | Not suitable for use with flammable gases | ||
| Self-Test Program | Automatic when power applied | ||
| Mode of Operation | Continuous | ||
| Pressure Sensor | |||
| Display Range | -50 to 300 mmHg | ||
| Display Resolution | 1 mmHg | ||
| Display Accuracy | ± 2 mmHg or ± 2% of reading (typical**) | ||
| Temperature Sensor | |||
| Display Range | 15 to 55°C | ||
| Display Resolution | 0.1°C | ||
| Oxygen Sensor | |||
| Display Range | 0 - 200 mmHg | ||
| Display Resolution | 1 mmHg | ||
| Display Accuracy | ± 1 mmHg | ||
| Analog output | |||
| Impedance | Apparent impedance is 0 ohms for loads <10 mA | ||
| Scale | 10 mV/mmHg; 10 mV/°C; 10 mV/°F | ||
| Bandwidth | Analog front end: -3dB @ 1.07 KHz | ||
| Conversion and linearization: 3277sps | |||
| Analog Output: -3dB @ 1.00 KHz | |||
| Digital output | |||
| RS232 | 19200 Baud, n, 8, 1, 200 samples/s | ||
| High speed SPI interface | 3277 samples/s | ||
| ^^ Blood pressure measurements in the range 0-300 mmHg can be made to 10,000 ft. | |||
| **AAMI requires measurement accuracy of ± 2 mmHg or ± 5% of reading for medical applications. | |||