| 1. | Place the Instrument in a well
ventilated location, attach the line cord to an appropriate 110/220 VAC
outlet, and switch Instrument power to ON (back panel). A green light (front panel) indicates Instrument is ON. |
||
| 2. | Insert the catheter optical plug into the instrument optical receptacle. Make sure the catheter optical plug is orientated properly during insertion. | optical receptacle | |
| 3. | Insert the catheter calibration key into the instrument calibration key receptacle. | calibration key receptacle | |
| 4. | Connect appropriate
cables to the analog or RS232 (digital) outputs located on the rear panel. |
analog output | |
| 5. | Wait for the Instrument
to initialize hardware/software and bring the optical components to operating
temperature. An LED display 'racetrack' indicates the Instrument is initializing and warming to operating temperature. |
racetrack pattern | |
| 6. | FOR PRESSURE SENSORS When the LED display shows "CAL - 0 - CAL - 0 etc.", the Instrument is ready for zero pressure setting:
If the LED display shows "Run",
the sensor is ready for use and the catheter may be placed at desired
measurment location. |
|
data enter key |
| 6. | FOR TEMPERATURE SENSORS
When the LED display shows "CAL - 40 - CAL - 40 etc.", insert temperature sensor into the front panel orifice labeled "40 C" and press the data enter key once. When calibration is finished the LED display will show "RUN"; the temperature sensor may now be removed and placed at the desired measurement location. |
data enter key |